Pulse Oximeter User Manual

Model: BM1000


1. Install Battery

  1. To remove the back cover compartment, push the white button and follow the direction of the printed arrows.

  2. Install two AAA batteries into the battery compartment. Match the plus (+) and minus (-) signs in the compartment. Incorrect polarity may damage the oximeter.

  3. Slide the battery door cover horizontally in the direction of the arrow.


2. Measurement

  1. Press the Function button.

  2. Place one finger into the rubber hole of the oximeter (insert fully) before releasing the clamp, with the nail upward.

  3. The data (SpO₂ and Pulse Rate) will display. Do not move the finger; remain motionless during reading.

  4. Once finished, “Thank you” will display. Remove finger. Device turns off automatically.

NOTE:

  • Do not use on damaged extremities or with arterial catheters.

  • Avoid simultaneous SpO₂ and NIBP measurements on the same arm.

  • Do not use for patients with pulse rates <30 bpm.

  • Cover sensor under strong light to avoid inaccurate readings.

  • Disinfect before using on different patients.

  • Change sensor location and check skin integrity every 2 hours.

Factors Affecting Accuracy:

  • Nail polish, artificial nails, dyes, or pigmented creams.

  • Shock, anemia, hypothermia, vasoconstrictive drugs.

  • Poor perfusion or incorrect sensor placement.


Product Description

  • Small, compact, reliable device for SpO₂ and pulse rate checks.

  • Components: mainboard, OLED display, dry batteries.

Intended Use:

  • Intermittent checks of oxygen saturation and pulse rate for adults in clinical environments.

  • Not for continuous monitoring.

Contraindications:

  • Not suitable for injured skin tissue.

Measurement Principle:

  • Based on spectrophotometry and Lambert Beer Law.

  • Uses two wavelengths: 666 nm (red) and 905 nm (infrared).


Function Description

  • Press POWER/FUNCTION to power on and rotate display.

  • Press and hold for 3 seconds to show IMEI and SIM card number.

  • Invalid value displayed when no finger is inserted.


Warnings

  • Keep away from children (choking hazard).

  • Do not use near flammable anesthetics or in MRI/CT environments.

  • No alarm function; not for long-term monitoring.

  • Do not modify device; maintenance only by authorized personnel.

  • Not waterproof; avoid liquids and mechanical damage.

  • Dispose per local regulations.

  • Use only AAA alkaline batteries.


Technical Specifications

  • Display: OLED

  • SpO₂ Range: 0–100% | Accuracy: ±3% (70–100%)

  • Pulse Rate Range: 25–250 bpm | Accuracy: ±2 bpm

  • Low Perfusion: 0.5–20%

  • Voltage: DC 2.2–3.4 V

  • Battery: Two 1.5 V AAA alkaline batteries

  • Power Consumption: <50 mA

  • Size: 58 × 34 × 30 mm | Weight: 50 g (with batteries)

  • Operating Temp: +5–40°C | Storage: -10–50°C

  • Humidity: 15–80% (operation), 10–90% (storage)

  • LED Wavelength: 666 nm / 905 nm | Output Power: <0.1 mW


Cleaning & Disinfection

  • Do not immerse in water.

  • Recommended disinfectants: 70% ethanol, 70% isopropanol, 2% glutaraldehyde.

  • Remove batteries before cleaning.

  • Wipe with soft cloth moistened with water or disinfectant; dry naturally.


Troubleshooting

Problem

Possible Reason

Solution

SpO₂/PR not displayed

Finger not deep enough; movement

Reposition finger; keep still

Device won’t power on

Batteries drained or incorrect installation

Replace/reinstall batteries

Screen turns off suddenly

Normal or low battery

Replace batteries

Unstable readings

Finger shaking or poor positioning

Keep calm; reposition finger


Electromagnetic Compliance

  • Complies with IEC 60601 standards and FCC Part 15 rules.

  • Suitable for domestic and clinical environments.